DIAGNOSTIC RADIOGRAPHY · NTA LEVEL 4 · SEMESTER TWO
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY ACCESSORY EQUIPMENT
CRT04208 · Radiology Quality Assurance
Study Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY ACCESSORY EQUIPMENT using the sections below. Use the topic navigation to continue through Radiology Quality Assurance.
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY & Imaging – ACCESSORY EQUIPMENT
2024/2025
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
AIM
To provide the knowledge and practical skills necessary to establish and carry out a quality control program for accessory equipment
OBJECTIVES
Understand and carry out quality control procedures, evaluation and recommend action
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
COLLIMATOR
- Visual inspection
- Accuracy of scales
- Change a light bulb
- Light beam / X-ray beam alignment and centring
Lead shutter efficiency
The collimator (light beam diaphragm), provides the radiographer with an easy to use , accurate, method of controlling the size of the x-ray field and placing it over the area of interest, thus reducing the radiation dose to the patient and improving the quality of an image The collimator receives much use and is vulnerable to knocks, often resulting in inaccuracy of the light beam / X-ray beam coincidence, blown light bulbs, electrical and mechanical problems
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY & Imaging
ACCESSORY EQUIPMENT
VISUAL INSPECTION
- Rotation.
- Stability
- Knobs for shutter control
- Accuracy of shutter setting
- Shutter chain drive
- Shutters
- Housing
- Window
- Light timer
- Timer switch
- Cables
- Plugs
- Light bulb
Light / x-ray beam alignment
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
ACCURACY OF THE SCALE TEST
The aperture size scales on the collimator used to set the aperture size, so it is advisable to make sure that the scales are accurate Inaccuracy of scales may lead to inaccurate collimation
FREQUENCY OF TEST
- 6 monthly
EQUIPMENT REQUIRED
- Collimator to be tested
- 100cm rule
2024/2025
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
ACCURACY OF THE SCALE TEST
METHOD
Set 100cm focus to tabletop distance Using the 100cm FFD scale on the collimator, set various aperture sizes with the light on and measure the resultant light areas at table top level Compare with setting on the collimator
EVALUTION
- Light area measured should be the same as the setting on collimator
ACTION
- If settings are not accurate and the solution is not a simple one, call an x-ray engineer
File a report
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY & Imaging
ACCESSORY EQUIPMENT
CHANGING A LIGHT BULB
A light bulb can fail at any time without waning. It is important then, to have the correct spare bulb readily available Staff should be capable of replacing a faulty bulb
FREQUENCY
- As necessary
EQUIPMENT
- Suitable light bulb
- Cloth or tissue
Screwdriver
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
CHANGING A LIGHT BULB
METHOD
- Switch off the power
- Remove relevant collimator housing
Check that spare is the correct type Before removing the old bulb, check to see if the replacement bulb must be fitted a specific way round Replace faulty bulb with the new one If the bulb is of the quartz type, ensure that it is not handed with bare fingers, as body oil on the bulb will shorten its life Replace housing and tighten screws
- Test
Ensure that the spare bulb is replaced as soon as possible
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST
The purpose of the light in a collimator is to allow more accurate collimation of the x-ray beam. The light must coincide with the x-ray beam. The light beam relies on the accurate positioning of the light bulb and angled mirror inside the collimator Should either the mirror or light bulb be dislodged, then errors in collimation could occur, resulting in areas of interest being excluded from the field, or too large an area being irradiated
FREQUENCY OF TEST
- 6 monthly
As necessary
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST
EQUIPMENT REQUIRED
One 24x 30 cm loaded cassette
Alignment test tool, commercially made test tools are available, but a simple alternative is to use eight coins or four paper clips Lead marker or ninth coin
METHOD
- Make sure that the table is level and the central ray at 90° to the tabletop
- Place a loaded cassette on the tabletop face up
- Set a FFD of 100cm
Switch on the collimator light
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST
METHOD
Centre to the middle of the cassette Collimate to within the edge of the cassette, leaving a 3cm border all round, that is outside the light field Place the coins in pairs, so that, where the coins touch, coincides with the edge of the light area All four borders of the light must be marked in this way Place the lead marker within one corner of the light field so that the film can be related to the light /x –ray field and hence the collimator shutters Make an exposure sufficient to blacken the film Process the film
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
GRID
The purpose of the grid is to reduce the amount of low energy photons reaching the detector. Although the quality of the image is improved, it does mean a significant increase in exposure, and therefore dose to the patient.
A grid looks like a simple thin sheet of soft metal, but it is in fact a precision made of equipment and easily damaged. It is made up of a large number of fine lead strips interspaced between radiolucent strips. If the lead strips become distorted the grid will be less efficient and irregular densities will be created on the film If the grid is used wrongly, abnormal densities ( grid cut off) will be produced e.g. a focus grid used up side down will create reduction in density the further the area is from the centre line
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
SPECIFICATIONS
The detail of the grid structure should be marked on the surface of the grid either by a label or imprinted into the metal GRID RATIO :- The ratio of the height of the lead strips to the distance between them GRID LINE:- The number of lead strips in a grid. Stated as the number of lead strips to the centimeter ( or to the inch ) FOCAL RANGE:- Grids are designed to be used within given focus film distances ( source image distances ) dependent on the way the lead strips have been focused GRID LINE DIRECTION :- The direction in which the grid lines travel is indicated by a line down the centre of the grid on the tube side. Cross – hatch grids have second line running at 90° to the first TUBE SIDE:- The tube side is shown by a label stating TUBE SIDE, or an x-ray tube symblol. The lead strips direction line also indicates the tube side
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
CARE OF THE GRID:-
Stationary grids should have an added, rigid, PVC protective cover When a protective cover is added the grid specifications must be recorded on the new cover for future reference Do not bend, drop or dent the grid
- Clean regularly
- Store in a safe place
USE OF A STATIONARY GRID
- Use a grid that is the same size as the cassette
- Place the grid against face of cassette with northing between it and the cassette face
- Check that the tube side of the grid is toward the tube
tched off when not in use and any safety lock keys removed
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
USE OF A STATIONARY GRID
- Use strips of adhesive tape to hold the grid in place
- Use the recommended focus film distance ( source image distance) for that grid
The x-ray beam must not be angled across grid lines
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
GRID LINE DAMAGE TEST
FREQUENCY OF TEST
- 6 monthly
- As necessary
EQUIPMENT REQUIRED
- Grid to be tested
- One loaded cassette, same size as grid
- One lead marker
METHOD
VISUAL INSPECTION
- Dents, bends, creases, damaged corners
- Specifications marked
Cleanliness
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
- X-RAY
- Place cassette on tabletop, face up
- Place grid to cover cassette, tube side up
- Set a FFD in the middle of the known focal range of the test grid or 100cm
- Centre to middle of grid
- Central ray at 90° to cassette / grid
- Collimate to cover cassette / grid
- Set an exposure of 50kV 10mAs
- Expose
Process film
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
EVALUTION
- Is grid line pattern regular ?
- Is overall density even ?
ACTION
If the grid line pattern or density is unacceptable repeat the test using FFDs of 10cm either side of that used in the first test If the images are still unacceptable, replace the grid File a report
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LEAD RUBBER APRONS AND GLOVES
Lead readily absorbs radiation and is used in many forms and thicknesses in the field of radiation protection. e.g. lead sheet, lead rubber, lead acrylic Lead rubber has advantage of being flexible and therefore be made into items which can be worn e.g. aprons and gloves.
Lead rubber tends to deteriorate over a period of time, with regular use and mishandling. Regular care, cleaning and testing are important
OTHER LEAD RUBBER ITEMS
- Gonad shields
- Thyroid shields
- Small sheets of differing size
Shielding on screening units
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LEAD RUBBER APRONS AND GLOVES
- CARE
- Weekly cleaning or as necessary. Clean with soap and water
- Never fold any item of lead rubber
- Aprons should be hung up using strong, specially designed hangers
- Store gloves flat or upright on a custom made holder
Do no store near a heat source
The lead equivalent ( effectiveness) of all forms of lead rubber should be known. Aprons for instance are designed only to protect against scatter. They have a lead equivalent of between 0.25mm and 0.5mm. Lead sheet may have a lead equivalent of 2.0mm or 3.0mm
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LEAD RUBBER APRONS AND GLOVES
VISUAL INSPECTION TEST
FREQUENCY OF TEST
- 6 monthly
- As necessary
EQUIPMENT REQUIRED
- Items to be tested
METHOD
- Check effectiveness and condition of fastenings on aprons ( if any)
- Inspect surface covering for tears or sign of deterioration
- Inspect all seams where appropriate
- Check apron hanger devices
Assess storage location
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LEAD RUBBER APRONS AND GLOVES
EVALUATION
- General condition and cleanliness
- Possibility of cracks
- Storage location and hanging devices
- Are staff using the apron hangers
ACTION
- Initiate any repairs or cleaning required
- Carry out x-ray check
- Relocate storage location if unsuitable
- Re-design hangers if unsuitable
- Re-educate staff if the are not hanging / storing items correctly
- Replace if necessary
File a report
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
TEST TO DETECT CRACKING OF APRONS OR GLOVES
FREQUENCY OF TEST
- Yearly
- As necessary
EQUIPMENT REQUIRED
Screening x-ray unit (if a screening unit is not available, use a general purpose unit and 35cm x 43cm cassettes) Items to tested
METHOD
- Spread article to be tested on the tabletop
- Use screening system or examine whole article.
Use different kVps
If a general purpose x-ray unit is to be used, place cassettes under test article and separate exposures. Ensure that the whole article is examined
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
TEST TO DETECT CRACKING OF APRONS OR GLOVES
EVALUTION
- Look for cracks or other abnormal variations of density in the image
ACTION
Items found to be defectives should be replaced Items being withdrawn can be cut up into smaller pieces, avoiding the cracked sections. These can be used for gonad shielding or shielding parts of the film File a report
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
A useful aid to maintaining the position of the patient. A range of pad sizes and shapes should be available in each x-ray room. They must be radiolucent Sets of film foam pads are available from manufacturers ( a polyethylene foam is commonly used ) pads may be improvised using other materials but they must be checked for radiolucency
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
ROUTINE CARE
- Care of positioning pads should be ongoing
- Try to minimize the risk of contamination by:
- Contrast media
- Plaster of paris
- Blood
- Urine
- Inspect and wash regularly
Do not use for any other purpose
It may be preferable to cover with waterproof radiolucent material, or removable cotton covers or both
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
POSITIONING PAD TESTS
- Visual
- X-ray
VISUAL TEST
FREQUENCY OF TEST
- Weekly
- As necessary
EQUIPMENT REQUIRED
- Pads to be inspected
INSPECT FOR:
- Likelihood of contamination
General cleanliness
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
INSPECT FOR:
- Shape
- Smell
- Particles breaking off
- Firmness
EVALUATION
- Unhygienic ?
- Likelihood of contamination by radio opaque substances ?
Shape, firmness, general deterioration ?
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
ACTION
Wash and re –test if necessary
If there is a likelihood of contamination with a radiopaque substance then carry out the x-ray test If the pad has deteriorated to the extent that it no longer fulfil its function effectively then replace it File a report
X-RAY TEST
FREQUENCY OF TEST
- 6 monthly
As necessary
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
EQUIPMENT REQURED
- X-ray unit
- Pads to be tested
- Loaded cassettes as necessary
- Lead marker
METHOD
Place a loaded cassette on the tabletop face up Place the pad on the cassette ( ensure that the cassette is big enough to cover whole pad) Set a FFD of 100cm
Collimate to cover the pad
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
METHOD
Place the marker in one corner of the x-ray field ( note which corner of the pad it relates to) Set an exposure of 50kV and 6mAs
- Expose
- Process the film
EVALUATION
Look for radio opaque shadows within the pad
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
PATIENT POSITIONING AIDS
ACTION
- If radiopaque shadows are detected, wash and dry the pad
- Re – test
If opacities are still seen, replace the pad Continued use of the pad may be considered , provided it is identified as containing radiopaque material and not used where it might show on the films File a report
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
OTHER POSITIONING AIDS
PATIENT POSITIONING AIDS
A range of material can be used to assist in positioning the patient in specific positions, some radiolucent and some not Care should always be taken to ensure that radiopaque positioning aids do not obscure the image SANDBAGS:- These are radiopaque. They should have waterproof or removable cotton or linen covers that must be washed regularly FLOUR BAGS:- Similar to sandbags but are radiolucent. They mould to different shapes more readily, but are not so heavy as sandbags
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
OTHER POSITIONING AIDS
COMPRESSION BAND:- If the band is linen, detach from the end fixing mechanisms and wash as necessary If the band is plastic, simply wipe over with a damp cloth. The ratchet and fixing mechanisms should be inspected and tested six monthly or as necessary.
The band should also be tested every six moths or as necessary, for radiopaque substance contamination WATER BOTTLE:- Empty fixer or developer bottles full of water are useful as cassette supports, or in weight bearing views of a/c joints. Water is relatively radio –opaque. Little care is required, simply wipe the outside of the bottle with a damp cloth as necessary. Ensure that the lid is on tightly and that the bottle does not leak
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
OTHER POSITIONING AIDS
- Polystyrene blocks used to pack equipment can be cut to required shapes
- Polystyrene chips used in packing can be sealed into suitable sized cotton or linen bags
- Cardboard boxes can be cut down to suitable sizes and shapes and re- stuck
- Wooden blocks can be cut to shape
Folded blankets can also be useful
N/B It should be remembered that any improvised positioning aid checked as to its radiolucency before use
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY & Imaging – ACCESSORY EQUIPMENT
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
ACCURACY OF THE SCALE TEST
The aperture size scales on the collimator used be used to set the aperture size, so it is advisable to make sure that the scales are accurate Inaccuracy of scales may lead to inaccurate collimation
FREQUENCY OF TEST
- 6 monthly
EQUIPMENT REQUIRED
- Collimator to be tested
100cm rule
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST
The purpose of the light in a collimator is to allow more accurate collimation of the x-ray beam. The light must, therefore, coincide with the x-ray beam. The light beam relies on the accurate positioning of the light bulb and angled mirror inside the collimator Should either the mirror or light bulb be dislodged, then errors in collimation could occur, resulting in areas of interest being excluded from the field, or too large an area being irradiated
FREQUENCY OF TEST
- 6 monthly
As necessary
EQUIPMENT REQUIRED
One 24x 30 cm loaded cassette
Alignment test tool, commercially made test tools are available, but a simple alternative is to use eight coins or four paper clips Lead marker or ninth coin
METHOD
- Make sure that the table is level and the central ray at 90° to the tabletop
- Place a loaded cassette on the tabletop face up
- Set a FFD of 100cm
- Switch on the collimator light
- Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
LIGHT BEAM / X-RAY BEAM ALIGNMENT TEST by:-
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
GRID
The purpose of the grid is to reduce the amount of scattered radiation reaching the film. Although the quality of the image is improved, it does mean a significant increase in exposure, and therefore dose to the patient.
A grid looks like a simple thin sheet of soft metal, but it is in fact a precision made of equipment and easily damaged. It is made up of a large number of fine lead strips interspaced between radiolucent strips. If the lead strips become distorted the grid will be less efficient and irregular densities will be created on the film If the grid is used wrongly, abnormal densities ( grid cut off) will be produced e.g. a focus grid used up side down will create reduction in density the further the area is from the centre line
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
GRID TYPES
STATIONARY
- A loose grid which can be placed directly over the face of the cassette
- The grid and cassette size must be the same
- The grid ratio is usually 6: 1 or 8:1
- Cassettes are available with a built in, or added grid
MOVING
A grid used in a potter Bucky system, which moves from side to side during the x-ray exposure, in order to defuse out the images of the lead strips The grid ratio is usually 10:1 or 12:1 The grid ratio should be 16: 1 if used for high kV work
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
PARALLEL GRID
- Lead strips run parallel to one another in one direction only
- There are two types of parallel line grid, focused and non- focused
Moving and stationary grids are all of parallel construction
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
CROSS – HATCH GRID
Two sets of lead strips superimposed and running at 90° to one another Generally a stationary grid, used only for relatively high kV work and when on tube angulation is necessary This design is used only in stationary grids
Quality Assurance and Quality Control IN DIAGNOSTIC RADIOGRAPHY
ACCESSORY EQUIPMENT
- GRID
FOCUSED GRID
Focused grid has lead strips progressively incline more inward, the further they are from the centre line, so that they are focused to a pre – determined point above the grid Thin lead strips interspaced between equally thin strips of radiolucent material Covered top and bottom with thin Aluminium sheet